Fluke - 5440B Programmable Voltage Calibrator

Fluke - 5440B Programmable Voltage Calibrator
Price: £2,750.00
Manufacturer: FLUKE

The 5440B direct voltage calibrator is equipped with bright displays for clear visibility at any viewing angle. The 5440B is easy to use because the 40-character, alphanumeric display provided in addition to the numeric output display, makes operating instructions and error messages clear and readable in engineering units and English messages instead of coded numerics.

For semi-automated testing, the 5440B lets a complete test sequence of up to 60 steps be stored in the internal Procedure Storage memory and recorded for later use in the interchangeable plug-in Procedure Storage modules. A test sequence is recalled and executed one step at a time with each touch of the NEXT STEP key.

The excellent accuracy specifications for the 5440B are easily maintained in any lab by performing the simple automatic calibration procedures, which store the calibration constants in non-volatile memory. No hardware adjustments are required.

  • Full accuracy over extended temperature range of 10°C
  • Completely automatic internal calibration—no mechanical adjustments
  • Simplified 5-minute external calibration to traceable standards
  • Procedure Storage module for single button test setup
  • Completely programmable through the GPIB/IEEE-488 interface
  • Serial printer output for calibration constants and instrument test results
  • Output voltage: 0 to 1100 V
  • Output current: 0 to 60 mA up to 22 V, except divided outputs: 0 to 25 mA up to 1100 V
  • Uncertainty specification: ±(PPM of Output + Microvolts) for 90 days
    • 0 V to 11 V: 2.0 +5 µV
    • 11 V to 22 V: 2.0 + 8 µV
    • 22 V to 275 V: 3.5 + 100 µV
    • 275 V to 1100 V: 3.5 + 400 µV
    • Divided output 0 V to 220 mV: 5 + 0.5 µV
    • Divided output 0.22 V to 2.2 V: 4.5 + 1.0 µV
  • Output stability: ±(PPM of Output + Microvolts) for 30 days. Specifications apply for initial stabilization of two hours, constant ambient temperature of ±1°C, constant line voltage, constant load, and measurement bandwidth of 0.1 Hz to 1 Hz
    • 0 V to 11 V: 0.5 +3 µV
    • 11 V to 22 V: 0.5 + 4.5 µV
    • 22 V to 275 V: 1.0 + 60 µV
    • 275 V to 1100 V: 1.0 + 300 µV
    • Divided output 0 mV to 220 mV: 2 + 0.3 µV
    • Divided output 0.22 V to 2.2 V: 2 + 0.7 µV
  • Resolution
    • 0 V to 11 V: 1 µV
    • 11 V to 22 V: 1 µV
    • 22 V to 275 V: 10 µV
    • 275 V to 1100 V: 100 µV
    • Divided output 0 V to 220 mV: 0.01 µV
    • Divided output 0.22 V to 2.2 V: 0.1 µV
  • Output noise at 10 Hz to 10 kHz bandwidth
    • 0 mV to 220 mV: 5 µV
    • 0.22 V to 2.2 V: 15 µV
    • 0 V to 11 V: 30 µV
    • 11 V to 22 V: 50 µV
    • 22 V to 275 V: 150 µV
    • 275 V to 1100 V: 300 µV
  • Line regulation: Changes <±0.1 ppm of range for ±10% change from nominal line voltage
  • Load regulation: <±0.1 ppm change of output for change from no-load to full-load or from full-load to no-load for output load impedances >80 Ω
  • Common Mode Rejection: >140 dB for frequencies from dc to 400 Hz
  • Shock: Eighteen 20G ½-sinewave shocks