Description
Features:
- 0.2% accuracy for R, L, and C measurements
- 1 kHz and 120 Hz test frequencies
- Dual displays
- Built-in Kelvin test fixture
- Series and parallel measurement selection
- Wide measurement ranges
Measurement Modes: Measures R series or parallel; L and Q series or parallel; C and D series or parallel. All measurement modes are pushbutton selectable
Measurement Speed: Greater than three measurements per second, unqualified
Test Frequencies: 1 kHz and 120 Hz. Pushbutton selectable
Displays: Two LED numerical displays with automatically positioned decimal points and illumination of units. For RLC, five digits (99999) and simultaneously for DQ, four digits (9999)
Applied Voltage: 0.3 V rms maximum
Ranges: Pushbutton selection with automatic front-panel guidance
Specifications:
| Parameter | Minimum Low Extension | Basic Ranges | Maximum High Extension |
| R: 120 Hz | 00.001 ohm | 2 ohm to 2 Mohm | 99.999 Mohm |
| R: 1 kHz | 00.001 ohm | 2 ohm to 2 Mohm | 9.9999 Mohm |
| L: 1 kHz | 0.0001 mH | 0.2 mH to 200 H | 999.99 H |
| L: 120 Hz | 00.001 mH | 2 mH to 2000 H | 9999.9 H |
| C: 1 kHz | 0.0001 nF | 0.2 nF to 200 µF | 999.99 µF |
| C: 120 Hz | 00.001 nF | 2 nF to 2000 µF | 99999 µF |
| D (with C) | .0001 to 9.999 | ||
| Q (with L) | 00.01 to 999.9 |