Description
The Agilent 85032B calibration kit contains precision 50 ohm type-N standards used to calibrate Agilent network analyzers for measurement of devices with 50 ohm type-N connectors. Precision phase-matched 7 mm to 50 ohm type-N adapters can be included for accurate measurements of non-insertable devices. Standards include fixed terminations, open circuits, and short circuits in both sexes. This kit is specified from dc to 6 GHz.
- Return loss – loads
- ≥49 dB (DC to ≤2 GHz)
- ≥46 dB (>2 to ≤3 GHz)
- ≥40 dB (> 3 to ≤6 GHz)
- Return loss – adapters (type-N to 7-mm): ≥30 dB (DC to ≤6 GHz)
- Female open: ±0.501° ±0.484°/GHz deviation from nominal (DC to ≤6 GHz)
- Female short: ±0.490° ±0.385°/GHz deviation from nominal (DC to ≤6 GHz)
- Male open: ±0.501° ±0.234°/GHz deviation from nominal (DC to ≤6 GHz)
- Male short: ±0.441° ±0.444°/GHz deviation from nominal (DC to ≤6 GHz)